Anglia Instruments Ltd

Fast Trigger Spectrometers

Fast Trigger Spectrometers

buy now

The AvaSpec-2048FT is a special version of the AvaSpec-2048 with USB1 platform, that can start integration time with nanoseconds delay after an external trigger.

The external trigger can be applied by TTL signal to a Digital-in port in the AvaSpec's DB-15 interface connector.

A TTL output port is available to trigger a pulsed laser and the nanoseconds delay before starting the integration time can be set in the software (AvaSoft-FULL required).

The option for -42ns negative delay makes it extremely useful for measuring pulsed lasers.

The TTL output port can be used for Laser Induced Breakdown Spectroscopy (LIBS) and fluorescence applications, where measurements need to start after the laser has been fired.

The AvaSpec-2048FT can be delivered in all configurations and has all features of the AvaSpec-2048.

The multi-channel AvaSpec-2048FT-8 allows you to do LIBS measurements in the 200-1000nm range with 0.1nm resolution.

The USB2 platform offers even greater flexibility and speed for the fast trigger applications. The AvaSpec-2048/3648-USB2 have a programmable laser delay and laser pulsewidth.

For the USB2 platform, multiple scans can be controlled by the software (AvaSoft-FULL).

The AvaSpec-2048FT is supplied as standard with AvaSoft-BASIC, a complete manual, USB interface cable and will run on USB power (-SPU option). However, a PS-12V/1.0A power supply can be ordered separately for mains power use if this is preferred. In this case, the user will then have both options available. AvaSoft-FULL and application specific software can be ordered separately from the AvaSoft software library.

For further information and assistance, please contact us.

 

Technical Details

 

Spectrometer version

Parameter

AvaSpec-2048FT

AvaSpec-2048-USB2

AvaSpec-3648-USB2

t1 laser delay

Fixed 1.3µs

1µs – 89sec

1µs – 89sec

Steps for t1

N.A.

21ns

21ns

t2  integration delay

-4ns – 2.7ms

-20ns – 89sec

-20ns – 89sec

Steps for t2

42ns

21ns

21ns

t3 pulsewidth

Fixed 15µs

21ns – 1ms

21ns – 1ms

Steps for t3

N.A.

21ns

21ns

tint integration time

2ms – 60sec

1.1ms – 10min

10µs – 10min

Steps for tint

1ms

10µs

10µs

Timing Diagram

Timing Diagram

Ordering Information

Part Number

Description

AvaSpec-2048FT

Fast Trigger Fibre Optic Spectrometer, 75mm AvaBench, 2048 pixel CCD detector, self powered USB1.1, RS-232 interface, includes AvaSoft-BASIC, USB cable. Please specify grating, wavelength range and options

AvaSpec-2048-USB2

Fibre Optic Spectrometer, 75mm AvaBench, 2048 pixel CCD detector, self powered high speed USB2, USB2/RS-232 interface, includes AvaSoft-BASIC, USB cable. Please specify grating, wavelength range and options

AvaSpec-3648-USB2

Fibre Optic Spectrometer, 75mm AvaBench, 3648 pixel CCD detector, self powered high speed USB2, USB2/RS-232 interface, includes AvaSoft-BASIC, USB cable. Please specify grating, wavelength range and options

Options

-SPU

Includes switch for self powered USB1 (external power supply required for RS-232 interface)

-SPU2

Includes switch for self powered USB2 (external power supply required for RS-232 interface)

BT

Bluetooth® interface for USB2 platform only, includes antenna

SDXXX

Internal xxx MB SD card for on-board data saving, for USB2 platform only

DUV

Deep UV detector coating, required for use <350nm/>150nm

DCL-UV/VIS

Detector collection lens to enhance sensitivity, Quartz, 200-1100nm

SLIT-xx

Slit size, please specify xx = 10, 25, 50, 100, 200 or 500µm

OSF-YYY

Order sorting filter for reduction of 2nd order effects, please specify yyy = 385, 475, 515, 550 or 590nm

OSC

Order sorting coating with 590nm longpass filter for gratings VA, BB and VB, for the AvaSpec-2048/3648

OSC-UA

Order sorting coating with 350nm and 590nm longpass filter for grating UA, for the AvaSpec-2048/3648

OSC-UB

Order sorting coating with 350nm and 590nm longpass filter for grating UB, for the AvaSpec-2048/3648