Anglia Instruments Ltd

Reflection Probes

Reflection Probes

The FCR-7xx200-2 is a standard reflection probe to get spectral information of diffuse or directly reflecting or backscattering materials.

Via a standard SMA905 connector, light from a light source is coupled into a fibre bundle consisting of 6 fibers and carried to the probe end. The surface will selectively reflect light back into a 7th fibre. This fibre transfers the data to the output SMA905 connector which can be coupled to a spectrometer or spectrum analyzer. A higher number of fibres are possible to couple more energy from the light source and to increase the signal level.

For measurements under 90º, the FCR-90-Option, a special adapter with a mirror under 45º was developed. This can easily be mounted onto the tip of the reflection probe. The FCR-COL is available to focus the measurement spot on an extended distance.

Technical Details

Fibres

7 or 19 fibres of 200 or 400µm core, 6 or 17 light fibres, 1or 2 read fibres. N.A. = 0.22. Standard length 2m, splitting point in the middle

Wavelength range

200-800nm (UV/ViS) or 350-2000nm (VIS/NIR)

Connectors

2x SMA905

Probe end

Stainless Steel cylinder, 50mm long x 6.35mm diameter

Tubing

The optical fibres are protected by a silicon inner tube and a flexible chrome plated brass outer tubing, OD 5mm. This tube also provides stress relief.

Temperature

-30°C - 100°C. (High temperature version FCR-7UV400-2-ME-HTX)

Bending

Minimum bend radius: Short tern (few seconds) 20mm, long term: 60mm

Technical Drawing

Technical Drawing

Ordering Information

FCR-7xx200-2-ME

Reflection Probe, 7x200µm fibres, 2m length, SMA terminations

FCR-7xx400-2-ME

Reflection Probe, 7x400µm fibres, 2m length, SMA terminations

FCR-7xx400-2-ME-HTX

Reflection Probe, 7x400µm fibres, 2m length, SMA terminations, Max. 800°C

FCR-19xx200-2-ME

Reflection Probe, 19x200µm fibres, 2m length, SMA terminations

FCR-90-Option

90° Reflection add-on reflector for use with all 6.35mm reflection probes

FCR-COL-UV/VIS

Adjustable UV/VIS/NIR Collimating/focusing lens for FCR probes

xx = specify wavelength: UV for UV/VIS (200-800nm), IR for VIS/NIR (350-2000nm)