TE-Cooled Spectrometers
The AvaSpec-2048TEC is a special version of the AvaSpec-2048, where the Sony 2048 CCD detector is mounted on a one-stage Peltier cooling device.
This Peltier cooling element can reduce the temperature of the CCD chip by ca. 30 °C, improving the dynamic range by at least a factor of 10. As an additional benefit from the cooling, the dark noise is reduced by a factor of 2-3.
The above features enable the AvaSpec-2048TEC to be implemented in low light-level applications, such as fluorescence and Raman measurements, and where integration times of more than 5 seconds are needed.
The AvaSpec-2048TEC can be delivered as a single or dual channel instrument and has all the standard options, gratings and specifications as the standard AvaSpec-2048.
The AvaSpec-2048TEC is built into a desktop housing, has a cooling fan to actively ventilate the heatsink of the Peltier cooling element and an internal power supply.
For further information and assistance, please contact us.
Technical Details
|
Temperature Cooled CCD |
DT = ca. -30°C vs ambient |
|
Time to Stabilise |
1-2 minutes |
|
Dynamic Range improvement for it> 5 seconds |
>Factor 10 |
|
Dark Noise improvement for it> 5 seconds |
Factor 2-3 |
|
Peltier Cooling internal power supply |
Ca. 3.0V, 4A |
|
External power supply |
100-240VAC, 30W |
|
Dimensions |
310x235x135mm (Desktop Housing) |
Ordering Information
|
Part Number |
Description |
|
AvaSpec-2048-TEC |
Thermo-Electric Cooled Fibre Optic Spectrometer, 75mm AvaBench, 2048 pixel TEC CCD detector, USB1.1/RS-232 interface, includes AvaSoft-BASIC and USB cable. Supplied in desktop housing. Please specify grating, wavelength range and options |
|
AvaSpec-2048-TEC-2 |
Dual Channel Thermo-Electric Cooled Fibre Optic Spectrometer, 2x 75mm AvaBench, 2048 pixel TEC CCD detectors, USB1.1/RS-232 interface, includes AvaSoft-BASIC and USB cable. Supplied in desktop housing. Please specify grating, wavelength range and options for both channels |
|
Options | |
|
DUV |
Deep UV detector coating, required for use <350nm/>150nm |
|
DCL-UV/VIS |
Detector collection lens to enhance sensitivity, Quartz, 200-1100nm |
|
SLIT-xx |
Slit size, please specify xx = 10, 25, 50, 100, 200 or 500µm |
|
OSF-YYY |
Order sorting filter for reduction of 2nd order effects, please specify yyy = 385, 475, 515, 550 or 590nm |
|
OSC |
Order sorting coating with 590nm longpass filter for gratings VA, BB and VB, for the AvaSpec-2048 |
|
OSC-UA |
Order sorting coating with 350nm and 590nm longpass filter for grating UA, for the AvaSpec-2048 |
|
OSC-UB |
Order sorting coating with 350nm and 590nm longpass filter for grating UB, for the AvaSpec-2048 |