Transmission Immersion Probe with Variable Pathlength

The transmission dip probe with variable path length can be used to obtain online and inline absorbance measurements in fluids.
The 1/2” Probe has an optical path length which can be set by the user in a range from 0.5mm - 20mm and is therefore extremely versatile for high or low absorption fluids. The probe end is simply dipped into the fluid to make the measurement.
Via a standard SMA905 connector, light is coupled into a fibre bundle consisting of 6 fibres and carried to the probe end, where it crosses a user-defined physical gap of 0.25 - 10mm and reflects against a diffuse white material. The back reflected light returns into a 7th fibre. This fibre transfers the data to the output SMA905 connector which can be coupled to a spectrometer or spectral analyser. At the probe end, the read fibre is placed in the middle of the lighting fibres for best efficiency of the probe. Both bundles are led through one piece of flexible stainless steel tubing. In a ferrule, they are split and run separately for 1m to reach the SMA905 connectors.
Technical Details
|
Fibres |
6x illumination fibres, 1 detection fibre, all 200µm, 2m length |
|
Wavelength range |
200-800nm (UV/VIS) or 350-2000nm (VIS/NIR) |
|
Connectors |
2x SMA905 |
|
Optical path |
0.25 – 10mm physical gap, pathlength 0.5-20mm |
|
Probe end |
Stainless Steel or PEEK cylinder, 140-160mm long x 12.7mm (1/2”) diameter. Waterproof. |
|
Tubing |
The optical fibres are protected by a silicon inner tube and a flexible chrome plated brass outer tubing. This tubing also provides stress relief. OD 5mm. As an option, a waterproof, steel reinforced silicon tubing can be provided. |
|
Temperature |
-30°C - 100°C. (High temperature available on request) |
|
Bending |
Minimum bend radius: Short term (few seconds) 20mm, long term: 60mm |
Technical Drawing
Ordering Information
|
FDP-7xx200-2-VAR |
Transmission Dip Probe in SS with variable tip length, 2m length |
|
FDP-7xx200-2-VAR-PK |
Transmission Dip Probe in PEEK for corrosive fluids with variable tip length, 2m length |
|
FDP-TIP-VAR |
Replacement SS tip for transmission dip probe with variable pathlength |
|
FDP-TIP-VAR-PK |
Replacement PEEK tip for transmission dip probe with variable pathlength |